STX500-BT
SPECIFICATIONS
STX500-ICSA | STX500-HCSA | STX500-DC2SA | STX500-DRSA | STX500-HRSA | ||
Application | Incubation | Humid Cooling | Dry Storage | Dry MultiRange | Humid MultiRange | |
Temperatue Range (HT) | °C | 33..50 (70) | 04..25 | 04..25 | 04..50 | 04..50 (70) |
Temperature Stability | °C | +/- 0.2 | +/- 0.5 | +/- 1 | +/- 1.5 | +/- 0.3 |
Relative Humidity | % | max. 95 | max. 90 | 04..50 | 02..50 | 90..95 |
Communication | RS232 V24, 9600 Baud, ASCII | |||||
User Software | OS | Win32 / Windows NT / Windows2000 | ||||
Facility Requirements | 100V / 50Hz; 115V / 60Hz; 230V / 50Hz; switchable | |||||
Power Consumption | W | 450 | 400 | 400 | 700 | 500 |
Package Specifications
Weights and Dimensions
Unit | incl. Cassettes |
Net Weight | Tare | Gross weight | Package | Dimensions, cm (LxWxH) |
STX500-ICSA | none | 246 kg | 147 kg | 393 kg | Crate | 99 x 91 x 222 |
STX500-DC2SA | none | 282 kg | 147 kg | 429 kg | Crate | 99 x 91 x 222 |
STX500-HCSA | none | 282 kg | 147 kg | 429 kg | Crate | 99 x 91 x 222 |
STX500-HRSA | none | 282 kg | 147 kg | 429 kg | Crate | 99 x 91 x 222 |
Handling Specifications
Mean Access Time: 20.5 sec
Mean Load Time: 17.8 sec
Mean Unload Time: 23.1 sec
Installation Specifications
STX500-HCSA | STX500-DC2SA | STX500-HRSA | STX500-DFSA | |
mass flow for 50 Hz / 60 Hz AC main power | ||||
in cubic meters per hour | 925 / 1030 m3/h | 925 / 1030 m3/4 | 925 / 1030 m3/h | 1865 / 1970 m3/h |
in cubic feet per minute | 544 / 606 cfm | 544 / 606 cfm | 544 / 606 cfm | 1098 / 1160 cfm |
B > mm | D > mm | F > mm | G > mm | T > mm | |
STX500-SA | 100 | ~ 700 | 300 | 0 | 0 1) |
- encounter clearance for service accessibility on BT models
IC models do not require distances "F" and "B"
L > mm | R > mm | |
STX500-SA | 10 1) | 10 |
- Recommended for better accessibility
Minimum distances to the sidewalls of the units must be maintained in order to allow accessibility to main user door and in order to avoid the main user door to interfere.