logo

STS15

SPECIFICATIONS

  STS15-IC STS15-HC STS15-DC2 STS15-DR STS15-HR
Application   Incubation Humid Cooling Dry Storage Dry MultiRange Humid MultiRange
Temperatue Range (HT) °C 33..50 (70) 04..25 04..25 04..50 04..50 (70)
Temperature Stability °C +/- 0.2 +/- 0.5 +/- 1 +/- 1.5 +/- 0.3
Relative Humidity % max. 95 max. 90 04..50 02..50 90..95
Communication   RS232 V24, 9600 Baud, ASCII
User Software OS Win32 / Windows NT / Windows2000
Facility Requirements   100V / 50Hz; 115V / 60Hz; 230V / 50Hz; Switchable
Power Consumption W 400 400 400 400 400
Package Specifications

Weights and Dimensions

Unit incl.
Cassettes
Net Weight Tare Gross weight Package Dimensions, cm
(LxWxH)
STS15-IC 10 170 kg 101 kg 259 kg Crate 99 x 92 x 160
STS15-DC2 10 195 kg 101 kg 284 kg Crate 99 x 92 x 160
STS15-HC 10 195 kg 101 kg 284 kg Crate 99 x 92 x 160
STS15-HR 10 195 kg 101 kg 284 kg Crate 99 x 92 x 160
STS15-DF 10 195 kg 101 kg 284 kg Crate 99 x 92 x 160
STS15-AR 10 195 kg 101 kg 284 kg Crate 99 x 92 x 160
Handling Specifications

Mean Access Time: 19.3 sec

Mean Load Time: 15.1 sec

Mean Unload Time: 12.4 sec

Installation Specifications
Airflow STX44-BT
  STS15-HC STS15-DC2 STS15-HR STS15-DF
mass flow for 50 Hz / 60 Hz AC main power
in cubic meters per hour 450 / 525 m3/h 450 / 525 m3/4 450 / 525 m3/h 925 / 1030 m3/h
in cubic feet per minute 265 / 309 cfm 265 / 309 cfm 265 / 309 cfm 544 / 606 cfm
STX Distances
  B > mm D > mm F > mm G > mm T > mm
STS15 100 ~ 700 300 0 0 1)
  1. encounter clearance for service accessibility on BT models

IC models do not require distances "F" and "B"

STX44-BT Distances
  L > mm R > mm
STS15 10 1) 10
  1. Recommended for better accessibility

Minimum distances to the sidewalls of the units must be maintained in order to allow accessibility to main user door and in order to avoid the main user door to interfere.