logo

STX1000 BT

SPECIFICATIONS

  STX1000-ICBT STX1000-HCBT STX1000-DC2BT STX1000-DRBT STX1000-HRBT
Application   Incubation Humid Cooling Dry Storage Dry MultiRange Humid MultiRange
Temperatue Range (HT) °C 33..50 (70) 04..25 04..25 04..50 04..50 (70)
Temperature Stability °C +/- 0.2 +/- 0.5 +/- 1 +/- 1.5 +/- 0.3
Relative Humidity % max. 95 max. 90 04..50 02..50 90..95
Communication   RS232 V24, 9600 Baud, ASCII
User Software OS Win32 / Windows NT / Windows2000
Facility Requirements   100V / 50Hz; 115V / 60Hz; 230V / 50Hz; switchable
Power Consumption W 1200 900 1300 700 1200
Package Specifications

Weights and Dimensions

Unit incl.
Cassettes
Net Weight Tare Gross weight Package Dimensions, cm
(LxWxH)
STX1000-ICBT none 435 kg 160 kg 595 kg Crate 145 x 103 x 225
STX1000-DC2BT none 565 kg 160 kg 725 kg Crate 145 x 103 x 225
STX1000-HCBT none 565 kg 160 kg 725 kg Crate 145 x 103 x 225
STX1000-HRBT none 565 kg 160 kg 725 kg Crate 145 x 103 x 225
STX1000-DF Bi-Level, 40 Cass. 666 kg 248 kg 914 kg Crate 158 x 141 x 223
Handling Specifications

Mean Access Time: 20.5 sec

Mean Load Time: 18.9 sec

Mean Unload Time: 22.1 sec

Installation Specifications
STX Distances
  B > mm D > mm F > mm G > mm T > mm
STX1000-BT 100 ~ 700 300 0 0
STX1000-DFBT 200 ~ 600 500 0 0

IC models do not require distances "F" and "B"

STX1000 Distances
  L > mm R > mm
STX1000 15 1) 15
  1. Recommended for better accessibility

Minimum distances to the sidewalls of the units must be maintained in order to allow accessibility to main user door and in order to avoid the main user door to interfere.