logo

STX500-BT

SPECIFICATIONS

  STX500-ICSA STX500-HCSA STX500-DC2SA STX500-DRSA STX500-HRSA
Application   Incubation Humid Cooling Dry Storage Dry MultiRange Humid MultiRange
Temperatue Range (HT) °C 33..50 (70) 04..25 04..25 04..50 04..50 (70)
Temperature Stability °C +/- 0.2 +/- 0.5 +/- 1 +/- 1.5 +/- 0.3
Relative Humidity % max. 95 max. 90 04..50 02..50 90..95
Communication   RS232 V24, 9600 Baud, ASCII
User Software OS Win32 / Windows NT / Windows2000
Facility Requirements   100V / 50Hz; 115V / 60Hz; 230V / 50Hz; switchable
Power Consumption W 450 400 400 700 500
Package Specifications

Weights and Dimensions

Unit incl.
Cassettes
Net Weight Tare Gross weight Package Dimensions, cm
(LxWxH)
STX500-ICSA none 246 kg 147 kg 393 kg Crate 99 x 91 x 222
STX500-DC2SA none 282 kg 147 kg 429 kg Crate 99 x 91 x 222
STX500-HCSA none 282 kg 147 kg 429 kg Crate 99 x 91 x 222
STX500-HRSA none 282 kg 147 kg 429 kg Crate 99 x 91 x 222
Handling Specifications

Mean Access Time: 20.5 sec

Mean Load Time: 17.8 sec

Mean Unload Time: 23.1 sec

Installation Specifications
Airflow STX44-BT
  STX500-HCSA STX500-DC2SA STX500-HRSA STX500-DFSA
mass flow for 50 Hz / 60 Hz AC main power
in cubic meters per hour 925 / 1030 m3/h 925 / 1030 m3/4 925 / 1030 m3/h 1865 / 1970 m3/h
in cubic feet per minute 544 / 606 cfm 544 / 606 cfm 544 / 606 cfm 1098 / 1160 cfm
STX Distances
  B > mm D > mm F > mm G > mm T > mm
STX500-SA 100 ~ 700 300 0 0 1)
  1. encounter clearance for service accessibility on BT models

IC models do not require distances "F" and "B"

STX500-SA Distances
  L > mm R > mm
STX500-SA 10 1) 10
  1. Recommended for better accessibility

Minimum distances to the sidewalls of the units must be maintained in order to allow accessibility to main user door and in order to avoid the main user door to interfere.