STX500-ICSA |
STX500-HCSA |
STX500-DC2SA |
STX500-DRSA |
STX500-HRSA |
Application |
|
Incubation |
Humid Cooling |
Dry Storage |
Dry MultiRange |
Humid MultiRange |
Temperatue Range (HT) |
°C |
33..50 (70) |
04..25 |
04..25 |
04..50 |
04..50 (70) |
Temperature Stability |
°C |
+/- 0.2 |
+/- 0.5 |
+/- 1 |
+/- 1.5 |
+/- 0.3 |
Relative Humidity |
% |
max. 95 |
max. 90 |
04..50 |
02..50 |
90..95 |
Communication |
|
RS232 V24, 9600 Baud, ASCII |
User Software |
OS |
Win32 / Windows NT / Windows2000 |
Facility Requirements |
|
100V / 50Hz; 115V / 60Hz; 230V / 50Hz; switchable |
Power Consumption |
W |
450 |
400 |
400 |
700 |
500 |
Package Specifications
Weights and Dimensions
Unit |
incl. Cassettes |
Net Weight |
Tare |
Gross weight |
Package |
Dimensions, cm (LxWxH) |
STX500-ICSA |
none |
246 kg |
147 kg |
393 kg |
Crate |
99 x 91 x 222 |
STX500-DC2SA |
none |
282 kg |
147 kg |
429 kg |
Crate |
99 x 91 x 222 |
STX500-HCSA |
none |
282 kg |
147 kg |
429 kg |
Crate |
99 x 91 x 222 |
STX500-HRSA |
none |
282 kg |
147 kg |
429 kg |
Crate |
99 x 91 x 222 |
Handling Specifications
Mean Access Time: 20.5 sec
Mean Load Time: 17.8 sec
Mean Unload Time: 23.1 sec
Installation Specifications
|
STX500-HCSA |
STX500-DC2SA |
STX500-HRSA |
STX500-DFSA |
mass flow for 50 Hz / 60 Hz AC main power |
in cubic meters per hour |
925 / 1030 m3/h |
925 / 1030 m3/4 |
925 / 1030 m3/h |
1865 / 1970 m3/h |
in cubic feet per minute |
544 / 606 cfm |
544 / 606 cfm |
544 / 606 cfm |
1098 / 1160 cfm |
|
B > mm |
D > mm |
F > mm |
G > mm |
T > mm |
STX500-SA |
100 |
~ 700 |
300 |
0 |
0 1) |
- encounter clearance for service accessibility on BT models
IC models do not require distances "F" and "B"
|
L > mm |
R > mm |
STX500-SA |
10 1) |
10 |
- Recommended for better accessibility