Description
Specification
Flyer CO2
Flyer N2
Flyer O2
Print
Gassing Option
Specifications
CO2
Measuring Range
0..10
% Vol.
Input Pressure
1
4
bar typ.
bar max.
Gas Flow
~10
l/min max.
Gas Consumption @ 5 % CO2, 37°C S
<1
~7
~8
l/h standby
l/h 2-min.-access cycles
l/h 30sec.-access cycles
Measuring Wavelength
4200
nm
Accuracy
+/- 3
% FS
Stability
+/- 5
% FS over 12 month
Repeatability
+/- 0.5
% at full scale
Response Time
~30
s
Warm-Up Time
5
30
min. operation
min. full accuracy
MTBF
5
years
N2
Measuring Range
75..99.9
% Vol.
Input Pressure
1
4
bar typ.
bar max.
Gas Flow
~10
l/min max
Accuracy
+/- 0.5
% FS
Stability
+/- 0.5
% FS over 12 month
Repeatability
+/- 1
% at full scale
Response Time
~30
s
Warm-Up Time
5
min. operation
MTBF
30000
hrs
O2
Measuring Range
0..25
% Vol.
Input Pressure
1
4
bar typ.
bar max.
Gas Flow
~10
l/min max.
Accuracy
+/- 0.5
% FS
Stability
+/- 0.5
% FS over 12 month
Repeatability
+/- 1
% at full scale
Response Time
~30
s
Warm-Up Time
5
min. operation
MTBF
30000
hrs