STX56-BT Flush Back
SPECIFICATIONS
STX56-ICBT Flush Back | STX56-HCBT Flush Back | STX56-DC2BT Flush Back | STX56-DRBT Flush Back | STX56-HRBT Flush Back | ||
Application | Incubation | Humid Cooling | Dry Storage | Dry MultiRange | Humid MultiRange | |
Temperatue Range (HT) | °C | 33..50 (70) | 04..25 | 04..25 | 04..50 | 04..50 (70) |
Temperature Stability | °C | +/- 0.2 | +/- 0.5 | +/- 1 | +/- 1.5 | +/- 0.3 |
Relative Humidity | % | max. 95 | max. 90 | 04..50 | 02..50 | 90..95 |
Communication | RS232 V24, 9600 Baud, ASCII | |||||
User Software | OS | Win32 / Windows NT / Windows2000 | ||||
Facility Requirements | 100V / 50Hz; 115V / 60Hz; 230V / 50Hz; switchable | |||||
Power Consumption | W | 450 | 400 | 400 | 700 | 500 |
Package Specifications
Weights and Dimensions
Unit | incl. Cassettes |
Net Weight | Tare | Gross weight | Package | Dimensions, cm (LxWxH) |
STX56-ICBT Flush Back | 2 | 88 kg | 52 kg | 140 kg | Crate | 83 x 79 x 148 |
STX56-DC2BT Flush Back | 2 | 125 kg | 52 kg | 177 kg | Crate | 83 x 79 x 148 |
STX56-HCBT Flush Back | 2 | 122 kg | 52 kg | 174 kg | Crate | 83 x 79 x 148 |
STX56-HRBT Flush Back | 2 | 125 kg | 52 kg | 177 kg | Crate | 83 x 79 x 148 |
STX56-DFBT Flush Back | 2 | 125 kg | 52 kg | 177 kg | Crate | 83 x 79 x 148 |
STX56-NCBT Flush Back | 2 | 89 kg | 52 kg | 141 kg | Crate | 83 x 79 x 148 |
Handling Specifications
Mean Access Time: 15.3 sec
Mean Load Time: 14.1 sec
Mean Unload Time: 16.4 sec
Installation Specifications
STX56-HCBT Flush Back | STX56-DCBT Flush Back | STX56-HRBT Flush Back | STX56-DFBT Flush Back | |
mass flow for 50 Hz / 60 Hz AC main power | ||||
in cubic meters per hour | 450 / 525 m3/h | 450 / 525 m3/4 | 450 / 525 m3/h | 925 / 1030 m3/h |
in cubic feet per minute | 265 / 309 cfm | 265 / 309 cfm | 265 / 309 cfm | 544 / 606 cfm |
B > mm | D > mm | F > mm | G > mm | T > mm | |
STX56-BT | 100 | ~ 500 | 100 | 0 | 0 1) |
- encounter clearance for service accessibility on BT models
IC models do not require distances "F" and "B"
L > mm | R > mm | |
STX56 BT / SA | 10 1) | 10 |
- Recommended for better accessibility
Minimum distances to the sidewalls of the units must be maintained in order to allow accessibility to main user door and in order to avoid the main user door to interfere.