Solution Image

STX56-BT

SPECIFICATIONS

  STX56-ICBT STX56-HCBT STX56-DC2BT STX56-DRBT STX56-HRBT
Application   Incubation Humid Cooling Dry Storage Dry MultiRange Humid MultiRange
Temperatue Range (HT) °C 33..50 (70) 04..25 04..25 04..50 04..50 (70)
Temperature Stability °C +/- 0.2 +/- 0.5 +/- 1 +/- 1.5 +/- 0.3
Relative Humidity % max. 95 max. 90 04..50 02..50 90..95
Communication   RS232 V24, 9600 Baud, ASCII
User Software OS Win32 / Windows NT / Windows2000
Facility Requirements   100V / 50Hz; 115V / 60Hz; 230V / 50Hz; switchable
Power Consumption W 450 400 400 700 500
Package Specifications

Weights and Dimensions

Unit incl.
Cassettes
Net Weight Tare Gross weight Package Dimensions, cm
(LxWxH)
STX56-ICBT 2 88 kg 52 kg 140 kg Crate 83 x 79 x 148
STX56-DC2BT 2 125 kg 52 kg 177 kg Crate 83 x 79 x 148
STX56-HCBT 2 122 kg 52 kg 174 kg Crate 83 x 79 x 148
STX56-HRBT 2 125 kg 52 kg 177 kg Crate 83 x 79 x 148
STX56-DFBT 2 125 kg 52 kg 177 kg Crate 83 x 79 x 148
STX56-NCBT 2 89 kg 52 kg 141 kg Crate 83 x 79 x 148
Handling Specifications

Mean Access Time: 15.3 sec

Mean Load Time: 14.1 sec

Mean Unload Time: 16.4 sec

Installation Specifications
Airflow STX44-BT
  STX56-HCBT STX56-DCBT STX56-HRBT STX56-DFBT
mass flow for 50 Hz / 60 Hz AC main power
in cubic meters per hour 450 / 525 m3/h 450 / 525 m3/4 450 / 525 m3/h 925 / 1030 m3/h
in cubic feet per minute 265 / 309 cfm 265 / 309 cfm 265 / 309 cfm 544 / 606 cfm
STX Distances
  B > mm D > mm F > mm G > mm T > mm
STX56-BT 100 ~ 500 100 0 0 1)
  1. encounter clearance for service accessibility on BT models

IC models do not require distances "F" and "B"

STX56-BT Distances
  L > mm R > mm
STX56 BT / SA 10 1) 10
  1. Recommended for better accessibility

Minimum distances to the sidewalls of the units must be maintained in order to allow accessibility to main user door and in order to avoid the main user door to interfere.

Top