STX280-SA
CLIMATE
| Type | Temperature | Relative Humidity | Remarks | Descr. | Order Nr. |
| STX280-ICSA | 33..50°C (70°C) | max. 95% | 1) 7) | Incubator | 9122 05 10 |
| STX280-HCSA | 04..25°C | max. 90% | 1) 5) | Humid Cooler | 9122 05 11 |
| STX280-DC2SA | 04..25°C | 04..50% | 1) 3) 5) | Dry Storage | 9123 05 07 |
| STX280-HRSA | 04..50°C (70°C) | 90..95% | 1) 5) | Humid Wide Range | 9122 05 12 |
| STX280-DR2SA | 04..50°C | 02..50% | 1) 3) 5) | Dry Wide Range | 9122 05 05 |
| STX280-ARSA | 04..50°C | 10..95% | 1) 2) 5) 6) | Humidity Controlled | 9122 05 21 |
| STX280-DFSA | -20..0°C | 02..45% | 1) 4) 5) | Deep Freezer | 9122 05 27 |
| STX280-NCSA | N/A | N/A | N/A | No Climate | 9122 05 14 |
| STX280-DHSA | 04..50°C 04..70°C |
2..30% 70..95% |
1) | Dry Humid | 9122 20 84 |
- Important information: To ensure optimal operation, it is necessary to specify all temperature and humidity set points in order
- Humidity is adjustable
- Adjustable dehumidification
- Up to +20°C on request
- Humidity depends on temperature setting
- MANDATORY REQUEST: All Temperature and Humidity operation set points must be specified
- The lower reachable operation temperature depends on the instrument's ambient temperature and its mechanical integration. It may also be affected when combined with Shaker or Waver options or special internal lift configurations
The climate information in the above table is of indicative nature only. For details, always contact Liconic Sales or Applications and have the information on your planned set points ready.