Solution Image

STX66-BT

SPECIFICATIONS

  STX66-ICBT STX66-IC2BT STX66-HCBT STX66-DC2BT STX66-DRBT STX66-HRBT STX66-HR2BT STX66-DF STX66-AR
Application   Incubation Incubation Humid Cooling Dry Cool Dry MultiRange Humid MultiRange Humid MultiRange Deep Freezing All Range**
Temperatue Range (HT) °C 33..50 (70) 33..50 (70) 04.0 04.0 04.01) .. 50.0 04.01) .. 50.0 (70) 04.01) .. 50.0 (70) -20.0 .. 0 04.01) .. 50.0
Temperature Stability °C +/- 0.05 +/- 0.05 +/- 0.4 +/- 0.2 1) +/- 0.4 / +/- 0.05 1) +/- 0.4 / +/- 0.05 1) +/- 0.4 / +/- 0.05 +/- 0.7 1) +/- 0.4 / +/- 0.05
Relative Humidity*) % < 95 50 .. 95 < 90 04 .. 25 02 .. 50 90 .. 95 50 .. 98 - 0.2 .. 98
Communication   RS232 V24, 9600 Baud, ASCII
User Software OS Windows 7/32 / Windows 7/64 / Windows 10
Supply Voltage
Fuse Rating
  100V / 50Hz; 115V / 60Hz; 230V / 50Hz
Power Consumption W 450 450 400 400 700 500 500 600 600

All Specifications at 22°C / 50% Ambient Conditions.
Non binding Data maybe typical Values. For more Accurate please contact our Applications.

*) May depend on Temperature setpoint. It is recommendet to communicate planned setpoint in the order.
**)Requires Specification of planned setpoints for Temperature and HT.


Package Specifications

Weights and Dimensions

Unit incl.
Cassettes
Net Weight Tare Gross weight Package Dimensions, cm
(LxWxH)
STX66-ICBT 2 88 kg 52 kg 140 kg Crate 83 x 79 x 148
STX66-DC2BT 2 125 kg 52 kg 177 kg Crate 83 x 79 x 148
STX66-HCBT 2 122 kg 52 kg 174 kg Crate 83 x 79 x 148
STX66-HRBT 2 125 kg 52 kg 177 kg Crate 83 x 79 x 148
STX66-DFBT 2 125 kg 52 kg 177 kg Crate 83 x 79 x 148
STX66-NCBT 2 89 kg 52 kg 141 kg Crate 83 x 79 x 148
Handling Specifications

Mean Access Time: 15.3 sec

Mean Load Time: 14.1 sec

Mean Unload Time: 16.4 sec

Installation Specifications
Airflow STX66-BT
  STX66-HCBT STX66-DCBT STX66-HRBT STX66-DFBT
mass flow for 50 Hz / 60 Hz AC main power
in cubic meters per hour 450 / 525 m3/h 450 / 525 m3/4 450 / 525 m3/h 925 / 1030 m3/h
in cubic feet per minute 265 / 309 cfm 265 / 309 cfm 265 / 309 cfm 544 / 606 cfm
STX Distances
  B > mm D > mm F > mm G > mm T > mm
STX66-BT 100 ~ 500 100 0 0 1)
  1. encounter clearance for service accessibility on BT models

IC models do not require distances "F" and "B"

STX44-BT Distances
  L > mm R > mm
STX66 BT / SA 10 1) 10
  1. Recommended for better accessibility

Minimum distances to the sidewalls of the units must be maintained in order to allow accessibility to main user door and in order to avoid the main user door to interfere.

Top