Solution Image

Climate

DESCRIPTION

Type Temperature Relative Humidity Remarks Descr.
IC 33..50°C (70°C) max. 95% 1) 7) Incubator
HC 04..25°C max. 90% 1) 5) Humid Cooler
DC2 04..25°C 04..50% 1) 3) 5) Dry Storage
HR 04..50°C (70°C) 90..95% 1) 5) Humid Wide Range
DR 04..50°C 02..50% 1) 3) 5) Dry Wide Range
AR 04..50°C 10..95% 1) 2) 5) 6) Humidity Controlled
DF -20..0°C 02..45% 1) 4) 5) Deep Freezer
NC N/A N/A N/A No Climate
DH 04..70°C 70..95% 1) Dry Humid
  1. Important information: To ensure optimal operation, it is necessary to specify all temperature and humidity set points in order
  2. Humidity is adjustable
  3. Adjustable dehumidification
  4. Up to +20°C on request
  5. Humidity depends on temperature setting
  6. MANDATORY REQUEST: All Temperature and Humidity operation set points must be specified
  7. The lower reachable operation temperature depends on the instrument's ambient temperature and its mechanical integration. It may also be affected when combined with Shaker or Waver options or special internal lift configurations

  8. The climate information in the above table is of indicative nature only. For details, always contact Liconic Sales or Applications and have the information on your planned set points ready.
Top