logo

STX56 SA

SPECIFICATIONS

  STX56-ICSA STX56-HCSA STX56-DC2SA STX56-DRSA STX56-HRSA
Application   Incubation Humid Cooling Dry Storage Dry MultiRange Humid MultiRange
Temperatue Range (HT) °C 33..50 (70) 04..25 04..25 04..50 04..50 (70)
Temperature Stability °C +/- 0.2 +/- 0.5 +/- 1 +/- 1.5 +/- 0.3
Relative Humidity % max. 95 max. 90 04..50 02..50 90..95
Communication   RS232 V24, 9600 Baud, ASCII
User Software OS Win32 / Windows NT / Windows2000
Facility Requirements   100V / 50Hz; 115V / 60Hz; 230V / 50Hz; Switchable
Power Consumption W 450 400 400 700 500
Package Specifications

Weights and Dimensions

Unit incl.
Cassettes
Net Weight Tare Gross weight Package Dimensions, cm
(LxWxH)
STX44-ICSA 2 88 kg 52 kg 140 kg Crate 83 x 79 x 148
STX44-DC2SA 2 125 kg 52 kg 177 kg Crate 83 x 79 x 148
STX44-HCSA 2 122 kg 52 kg 174 kg Crate 83 x 79 x 148
STX44-HRSA 2 125 kg 52 kg 177 kg Crate 83 x 79 x 148
STX44-DFSA 2 125 kg 52 kg 177 kg Crate 83 x 79 x 148
STX44-NCSA 2 89 kg 52 kg 141 kg Crate 83 x 79 x 148
Handling Specifications

Mean Access Time: 15.3 sec

Mean Load Time: 14.1 sec

Mean Unload Time: 16.4 sec

Installation Specifications
Airflow STX44-SA
  STX44-HCSA STX44-DCSA STX44-HRSA STX44-DFSA
mass flow for 50 Hz / 60 Hz AC main power
in cubic meters per hour 450 / 525 m3/h 450 / 525 m3/4 450 / 525 m3/h 925 / 1030 m3/h
in cubic feet per minute 265 / 309 cfm 265 / 309 cfm 265 / 309 cfm 544 / 606 cfm
stx-distances
  B > mm D > mm F > mm G > mm T > mm
STX44-SA 300 ~ 500 300 0 2) 0 1)
  1. encounter clearance for service accessibility on BT models
  2. Do not remove Subframe

IC models do not require distances "F" and "B"

STX44-SA Distance
  L > mm R > mm
STX44 BT / SA 10 1) 10
  1. Recommended for better accessibility

Minimum distances to the sidewalls of the units must be maintained in order to allow accessibility to main user door and in order to avoid the main user door to interfere.