STF130
CLIMATE
| Type |
Temperature |
Relative Humidity |
Remarks |
Descr. |
Order Nr. |
| STF130-IC |
33..50°C (70°C) |
max. 95% |
1) 7) |
Incubator |
9141 01 96 |
| STF130-HC |
04..25°C |
max. 90% |
1) 5) |
Humid Cooler |
9141 01 97 |
| STF130-DC2 |
04..25°C |
04..50% |
1) 3) 5) |
Dry Storage |
9141 02 28 |
| STF130-HR |
04..50°C (70°C) |
90..95% |
1) 3) 5) |
Humid Wide Range |
9141 01 98 |
| STF130-DR |
04..50°C |
02..50% |
1) 3) 5) |
Dry Wide Range |
9141 02 15 |
| STF130-AR |
04..50°C |
10..95% |
1) 2) 5) 6) |
Humidity Controlled |
9141 02 16 |
| STF130-DF |
-20..0°C |
02..45% |
1) 4) 5) |
Deep Freezer |
9141 02 00 |
| STF130-NC |
N/A |
N/A |
N/A |
No Climate |
9141 02 29 |
- Important information: To ensure optimal operation, it is necessary to specify all temperature and humidity set points in order
- Humidity is adjustable
- Adjustable dehumidification
- Up to +20°C on request
- Humidity depends on temperature setting
- MANDATORY REQUEST: All Temperature and Humidity operation set points must be specified
- The lower reachable operation temperature depends on the instrument's ambient temperature and its mechanical integration. It may also be affected when combined with Shaker or Waver options or special internal lift configurations
The climate information in the above table is of indicative nature only. For details, always contact Liconic Sales or Applications and have the information on your planned set points ready.