logo

STX220-BT MR

CLIMATE

Type Temperature Relative Humidity Remarks Descr. Order Nr.
STX220-ICBT MR 33..50°C (70°C) max. 95% 1) 7) Incubator 9123 05 10 +9122 05 17
STX220-HCBT MR 04..25°C max. 90% 1) 5) Humid Cooler 9123 05 11 +9122 05 17
STX220-DC2BT MR 04..25°C 04..50% 1) 3) 5) Dry Storage 9123 05 08 +9122 05 17
STX220-HRBT MR 04..50°C (70°C) 90..95% 1) 5) Humid Wide Range 9123 05 12 +9122 05 17
STX220-DRBT MR 04..50°C 02..50% 1) 3) 5) Dry Wide Range 9122 05 22 +9122 05 17
STX220-ARBT MR 04..50°C 10..95% 1) 2) 5) 6) Humidity Controlled 9123 05 19 +122 05 17
STX220-DFBT MR -20..0°C 02..45% 1) 4) 5) Deep Freezer 9122 05 31 +9122 05 17
STX220-NCBT MR N/A N/A N/A No Climate 9123 05 15 +9122 05 17
  1. Important information: To ensure optimal operation, it is necessary to specify all temperature and humidity set points in order
  2. Humidity is adjustable
  3. Adjustable dehumidification
  4. Up to +20°C on request
  5. Humidity depends on temperature setting
  6. MANDATORY REQUEST: All Temperature and Humidity operation set points must be specified
  7. The lower reachable operation temperature depends on the instrument's ambient temperature and its mechanical integration. It may also be affected when combined with Shaker or Waver options or special internal lift configurations

  8. The climate information in the above table is of indicative nature only. For details, always contact Liconic Sales or Applications and have the information on your planned set points ready.