logo

STX44-MIBT

CLIMATE

Type Temperature Relative Humidity Remarks Descr. Order Nr.
STX44-ICMIBT 33..50°C (70°C) max. 95% 1) 7) Incubator 9118 12 93
STX44-HCMIBT 04..25°C max. 90% 1) 5) Humid Cooler 9118 12 94
STX44-DC2MIBT 04..25°C 04..50% 1) 3) 5) Dry Storage 9118 12 95
STX44-HRMIBT 04..50°C (70°C) 90..95% 1) 5) Humid Wide Range 9118 13 00
STX44-DRMIBT 04..50°C 02..50% 1) 3) 5) Dry Wide Range 9118 12 96
STX44-ARMIBT 04..50°C 10..95% 1) 2) 5) 6) Humidity Controlled 9118 12 97
STX44-DFMIBT -20..0°C 02..45% 1) 4) 5) Deep Freezer 9118 12 98
STX44-NCMIBT N/A N/A N/A No Climate 9118 12 99
  1. Important information: To ensure optimal operation, it is necessary to specify all temperature and humidity set points in order
  2. Humidity is adjustable
  3. Adjustable dehumidification
  4. Up to +20°C on request
  5. Humidity depends on temperature setting
  6. MANDATORY REQUEST: All Temperature and Humidity operation set points must be specified
  7. The lower reachable operation temperature depends on the instrument's ambient temperature and its mechanical integration. It may also be affected when combined with Shaker or Waver options or special internal lift configurations

  8. The climate information in the above table is of indicative nature only. For details, always contact Liconic Sales or Applications and have the information on your planned set points ready.